Total Phosphorous-Nitrogen analyzer is a continuous monitoring system that adopts manual analysis method for monitoring total nitrogen and total phosphorous values present in the sample. Total phosphorus and total nitrogen values are measured simultaneously every hour.
(30 minutes heating at 120℃ in the autoclave with potassium peroxodisulfate)
【Specification】
● Measurement method: Potassium peroxodisulfate decomposition
autoclave method (120℃, 30 minutes)
TP:Molybdenum blue (ascorbic acid) absorption spectroscopy
TN:UV absorption spectroscopy
● Measurement range :
TP: 0~0.1…20 mg/L (one range)
TN: 0~2.0…100 mg/L (one range)
(However, TN range≤50×TP range)
● Repeatability:±2.0%FS
(TN 0~50 mg/L , TP 0~10 mg/L)
● Measurement interval:Schedule function(Repeat/arbitrary
setting)
● Display:5.7 inch LCD touch panel
● Dimensions:700(W)×600(D)×1600(H) mm
<TPN-580R reduced reagent type spec.>
・Pottasium peroxodisulfate solution (1/2 of TPN-580)
・Sodium hydroxide solution (1/2 of TPN-580)
・Other necessary reagent quantities are same when compared to
standard TPN-580 analyzer.