AUTOMATIC MERCURY ANALYZER (HGM-108)

Ideal for process control !

HGM-108 analyzer uses cold vapor atomic absorption spectrophotometry for measuring inorganic mercury present in the sample. The standard measurement range of the analyzer is from 0-20μg/L, and the lowest is from 0.5μg/L. Continuous, automatic, accurate and stable measurements of the analyzer have replaced all the conventional methods in measuring inorganic mercury.

【Specification】
● Measurement method:Reducing vaporization-atomic
   absorptiometry
● Measurement range:0~10…20 μg/L
● Repeatability:±5%FS (by standard solution)
● Measurement interval:Internal start(JUST,0.5H,1H,2H,REPEAT)
  (External start- optional)
● Display
   1) Digital display
   ① Process and process time (STEP No.)
   ② Concentration (CONC.)
   ③ Range (RANGE)
   ④ Concentration alarm (C.ALARM)
   ⑤ Time (CLOCK)
   2) Display:LED display
   ① Control mode (START,LOAD,PAUSE)
   ② Operation mode (MEAS,CLEAN,CAL)
   ③ Alarms (C.ALARM,POWER,WATER,TROUBLE etc.)
   ④ Digital display selection
● Dimensions:800(W)×650(D)×1650(H) mm